Ramp Pressure; Humidifier - Breas Vivo 2 User Manual

Hide thumbs Also See for Vivo 2:
Table of Contents

Advertisement

Property
Setting resolution
Available in modes
4.2.3

Ramp Pressure

The ramp pressure setting is used for defining the start pressure when Ramp Up is
used:
IPAP shall start 2 cmH
Pressure is lower than EPAP
CPAP mode:
Property
Value range
Default value
Setting resolution
Available in modes
4.2.4

Humidifier

This setting defines whether a humidifier shall be used.
By default, this setting can be modified in Clinical mode only, but the Vivo 2 may be
configured to allow modification in home mode as well.
Doc. 007229 Y-2
Description
10 min
• PCV+A
• PSV
• S
• S/T
• T
• CPAP
O above Ramp Pressure, or 2 cmH
2
Description
• 2 cmH
O to the set value for EPAP (PSV and PCV+A
2
modes)
• 2 cmH
O to the set value for CPAP (CPAP mode)
2
• The set value for EPAP (PSV and PCV+A modes)
• The set value for CPAP (CPAP mode)
Below 10 cmH
O: 0.5 cmH
2
Above 10 cmH
O: 1.0 cmH
2
• PCV+A
• PSV
• S
• S/T
• T
• CPAP
O above EPAP if Ramp
2
O
2
O
2
Treatment Functions and Settings 47