Using The Remote Alarm Unit - Breas Vivo 2 User Manual

Hide thumbs Also See for Vivo 2:
Table of Contents

Advertisement

Supplemental Oxygen Supply Requirements
Property
Maximum flow
Maximum pressure
Supply source
Connector
Connect the Oxygen supply
1
Connect the oxygen adapter (article no. 005032) to the oxygen supply's tube.
2 Connect the oxygen adapter to the oxygen port at the back of the Vivo 2. See 3.3
Ventilator Back , page 30 for detailed information.
3 If using a device for monitoring the oxygen saturation, connect it according to the
supplier instructions.
6.6.5

Using the Remote Alarm Unit

Information about safety, warnings, product description, installation, usage, cleaning,
maintenance and technical specifications can be found in the user instruction for
Remote Alarm.
The Remote Alarm enables care providers and clinical personnel to monitor the Vivo 2
alarms remotely. The Remote Alarm forwards alarms from the Vivo 2.
1
Connect the Accessory box to the Communication port at the back of the Vivo 2.
2 Connect the remote alarm cable to the Nurse call/ Remote alarm port on the acces-
sory box.
Doc. 007229 Y-2
Requirement
30 l/min.
100 kPa
Source equipped with rotameter. Exam-
ples of supply sources:
• Oxygen cylinder
• Central oxygen supply
• Oxygen concentrator
The oxygen source shall be equipped
with the Breas Low Pressure Oxygen
Adapter, art. no. 005032
How to Use the Vivo 2 69
Vivo 2 User's Manual