Barco Demetra User Manual page 45

Hide thumbs Also See for Demetra:
Table of Contents

Advertisement

Image review
To have sufficient image quality, a suitable display should be used to review images in the Demetra Web
Application (see the part 'Requirements' in the chapter "Activation", page 12 for more details).
High magnetic environment
The device shall not be used in the high magnetic environment of an MRI scanner.
The installer shall assess the magnetic environment before installation or use of the device.
Water and moisture
Never expose the device to rain or moisture.
Never use the device near water - e.g. near a bathtub, washbasin, swimming pool, kitchen sink, laundry tub or
in a wet basement.
Eye safety
Hold your Demetra Scope so that the LEDs are aimed in the direction of the lesion to be examined.
Do not look directly into the bright LED light. It is recommended that patients close their eyes during
examinations.
The Demetra Scope is not intended to be used for lesions in the eyes.
Do not stare at operating lamp. May be harmful to the eyes. Product tested against IEC62471.
Glass temperature
The temperature of the contact glass can reach 46.5 °C.
Please verify the temperature of the contact glass and the condition of the patient before using contact mode.
Limit the contact duration to less than 1 minute, or use non-contact mode instead.
Breached or compromised surface
Avoid contact between the Demetra Scope and breached or compromised surfaces, even when the non-
contact cone is installed.
In case of large breached or compromised lesions, do not use the non-contact cone. Make sure to keep
appropriate distance between the Demetra Scope head and the lesion.
Risk patients
As the Demetra Scope contains several magnets (for the attachment of the non-contact cone and for the
installation in the battery charging station), pay attention when examining specific risk patients (e.g. patients
with a pacemaker, ...).
Malfunctions
Contact the Barco helpdesk under the following conditions:
If the power cord or plug is damaged or frayed.
If liquid has been spilled into the equipment.
If the equipment has been exposed to rain or water.
If the equipment does not operate normally when the operating instructions are followed. Adjust only those
controls that are covered by the operating instructions since improper adjustment of other controls may
result in damage and will often require extensive work by a qualified technician to restore the equipment to
normal operation.
If the equipment has been dropped or has been damaged.
If the equipment exhibits a distinct change in performance, indicating a need for service.
Technical data
For detailed information, see "Technical specifications", page 56.
Important information
K5902200 /02AC
Barco Demetra
45