• an external power supply If you ordered a Barco display controller, it’s also in the box together with its accessories. A dedicated user guide is available on the system CD. Keep your original packaging. It is designed for this display and is the ideal protection during transport and storage.
2. Parts, controls and connectors Display rear view Overview Image 2-2 Rear view +24 VDC power input +5 VDC, 0.5 A power output DVI 2 video input DisplayPort 2 video input Connector compartment cover Neck cover DisplayPort 1 video input Display stand cover DVI 1 video input Ethernet connector (see note below)
Prior to installing your Coronis Fusion 6MP and connecting all necessary cables, make sure to have a suitable display controller physically installed in your computer. If you are using a Barco display controller, please consult the user guide delivered with it to do this.
3. Display installation 2. Slide the cover upwards while holding the clips pressed. Unlocking the height mechanism In the factory, the height-positioning system in the display stand is locked with a red hook to prevent damage during transportation. You’ll have to remove this hook before adjusting your display height position.
3. Display installation 3. Keep the clip in the dedicated hole in case the display needs to be shipped later. To retain the hook for possible future transportation, insert the short, red end of the hook back into the stand of your display. Adjusting the display position Now that the height-positioning system of your display is unlocked, you can adjust the display position.
3. Display installation Do not try to pivot your display when attached to the stand. Trying to do so AUTION could cause serious damage to your display and its stand. Connecting the signal cables To get access to the connectors, remove the connector compartment cover. See "Re- moving the covers", page 7 .
3. Display installation Connecting the power cable To connect the power cable to the display: 1. Connect the supplied external DC power supply to the +24 VDC power input of your Coronis Fusion 6MP display. 2. Plug the other end of the external DC power supply into a grounded power outlet by means of the proper power cord delivered in the packaging.
3. Display installation Pay attention that all cables stay in the cable channel while re-attaching the cover. To re-attach the connector compartment cover 1. Slide the cover’s top in position and then push the cover’s bottom. You’ll hear a “click” sound of the cover’s clips when the connector compartment cover is in position.
Page 15
3. Display installation 1. Gently pull open both lower sides of the cover. 2. Slide the cover upwards while holding the lower sides pulled open. 3. Lift the plastic frame that covers the fixation of the panel to the stand. Turn it for 45 degrees to uncover the fixation screws.
Your Coronis Fusion 6MP display will be running in a basic video mode at a default re- fresh rate when first time starting up. If you are using a Barco display controller, please consult the dedicated user guide available on the system CD to install the drivers, soft- ware and documentation.
LCD displays use technology based on pixels. As a normal tolerance in the manufacturing of the LCD, a limited number of these pixels may remain either dark or permanently lit, without affecting the diagnostic performance of the product. To ensure optimal product quality, Barco applies strict selection criteria for its LCD panels.
The ’MediCal QAWeb’ system offers online service for high-grade Quality Assurance, providing maximum diagnostic confidence and uptime. Barco recommends to install MediCal QAWeb Agent and apply the default QAWeb policy at least. This policy includes calibration on regular intervals. Connecting to MediCal QAWeb Server offers even more possibilities.
4. Daily operation Switching on the display while it is in stand-by mode or vice versa can both be done by using the keys or by using the control wheel. To switch your display on using the keys: 1. Light up the keys indicator lights as previously described. 2.
4. Daily operation How to bring up the OSD menus using the control wheel 1. If not already done so, switch on the display as previously described. 2. Shortly turn the control wheel. As a result, the OSD main menu comes up in the bottom right corner of the screen. If no further actions are taken within the following 90 seconds however, the OSD menu will disappear again.
5. Advanced operation 5. ADVANCED OPERATION OSD menu position About the OSD menu position By default, the OSD menu comes up in the bottom center position of the screen. This position can be modified however. There’s choice from a number of predefined positions. To change the position of the OSD menu: 1.
5. Advanced operation Power status indicator light About the power status indicator light By default, when the display is switched on, the power status indicator light is dimmed. This behavior can be changed so that the power status indicator light will be blue when the display is switched on. When the display is in stand-by mode, the power status indicator light will always turn orange, even when the power status indicator light is disabled.
By enabling the Ethernet over USB, you allow the internal Coronis Fusion 6MP processor to communicate directly over USB with the connected PC. Enabling the Ethernet over USB is only recommended when you are using a non-Barco display controller. If you are using a Barco display controller, this communication will automatically be done over the connected video cable(s).
5. Advanced operation Enabling the USB endpoint is only recommended when you are using a non-Barco dis- play controller. If you are using a Barco display controller, this communication will au- tomatically be done over the connected video cable(s). To enable/disable USB: 1.
5. Advanced operation To enable/disable hibernation on your display: 1. Bring up the OSD main menu. 2. Navigate to the Configuration > Power Management menu. 3. Enter the Hibernate submenu. 4. Select Enabled/Disabled as desired and confirm. To specify the hibernate timeout: 1.
5. Advanced operation 3. Enter the Color Presets submenu. 4. Select one of the available Color Presets and confirm. 5.14 Color temperature About color temperature: It is possible to change the color temperature of your display. Color temperature can only be changed on your display when color presets is set to user.
In short, the DICOM display function results in more visible grayscales in the images. Barco recommends selecting the DICOM display function for most medical viewing applications. K5902055 CORONIS FUSION 6MP 14/11/2014...
5. Advanced operation User This display function will be automatically selected when display functions are defined by MediCal QAWeb. Gamma 1.8 Select one of these display functions in case the display is to replace a CRT display with a gamma Gamma 2.2 of 1.8 or 2.2 respectively.
5. Advanced operation The available reading rooms for your Coronis Fusion 6MP are: CR/DR/ MAMMO Corresponds to light conditions in diagnostic reading rooms for computed radiology, digital radiology or mammography. This setting has the lowest maximum ambient light. CT/MR/NM Corresponds to light conditions in diagnostic reading rooms for computed tomography, magnetic resonance or nuclear medicine scans.
Embedded QA is not a replacement for the Barco MediCal QAWeb solution. Although Embedded QA is a reliable option to perform a simple calibration or compliance test, Barco still highly recommends MediCal QAWeb as the solution of choice for calibration and QA. Medical QAWeb brings many benefits such as centralized asset management, the ability to schedule tasks, remote man-...
5. Advanced operation Current DICOM Settings Display function Shows the current display function. Ambient Light Compensation Shows the ambient light compensation status. Reading room Shows the selected reading room. To retrieve the DICOM status report: 1. Bring up the OSD main menu. 2.
5. Advanced operation 5.21.6 DICOM error threshold About DICOM error threshold The threshold to define the DICOM compliance can be modified in steps of 5% starting from 5 to 30%. When the maximum deviation is not bigger than the selected threshold, the compliance check will be OK. To set the DICOM error threshold: 1.
5. Advanced operation Automatic In this mode, your display automatically detects the connected video input signals, attaches them to the correct display side and applies the correct video settings to it (resolution, video encoding mode, refresh rate,...). No video settings are available when this mode is selected.
HDR1). This mode is usually used on color displays. Dual 10 (only with Barco display controller) Dual 10 is a Barco specific 10-bit video encoding mode which is usually used on grayscale displays and which allows full refresh rate.
5. Advanced operation About grayscale conversion modes Grayscale conversion modes specify how color generated on the display controller is converted to grayscale in your display. The available grayscale conversion modes are: No Conversion Use Red Channel This mode is intended for grayscale displays where gray is sent over the red channel.
5. Advanced operation 5.28 Display info About display info Your display serial number, color type, native resolution, firmware versions, etc. are available in a dedi- cated submenu of the OSD menu. To retrieve info about your display: 1. Bring up the OSD main menu. 2.
6. Cleaning your display 6. CLEANING YOUR DISPLAY Cleaning instructions To clean the display Clean the display using a sponge, cleaning cloth or soft tissue, lightly moistened with a recognized clean- ing product for medical equipment. Read and follow all label instructions on the cleaning product. In case of doubt about a certain cleaning product, use plain water.
7. Repackaging instructions Repacking your display How to repack your display 1. Place the empty box on a stable surface. 2. Slide the system CD, documentation CD and user guide in the dedicated bag attached at the inside of the box. 3.
8. Important information 8. IMPORTANT INFORMATION Safety information General recommendations Read the safety and operating instructions before operating the device. Retain safety and operating instructions for future reference. Adhere to all warnings on the device and in the operating instructions manual. Follow all instructions for operation and use.
8. Important information To fully disengage the power to the device, please disconnect the power cord from the AC inlet. Power cords: • Utilize a UL-listed detachable power cord, 3-wire, type SJ or equivalent, 18 AWG min., rated 250 V min., provided with a hospital-grade type plug 5-15P configuration for 120V application, or 6-15P for 240V application.
Page 43
(Also called RoHS of Chinese Mainland), the table below lists the names and contents of toxic and/or hazardous substances that Barco’s product may contain. The RoHS of Chinese Mainland is included in the MCV standard of the Ministry of Information Industry of China, in the section “Limit Requirements of toxic substances in Electronic Information Products”.
Page 44
“Electronic Information Products Pollution Control Labeling Standard” of Chinese Mainland, marked with the Environmental Friendly Use Period (EFUP) logo. The number inside the EFUP logo that Barco uses (please refer to the photo) is based on the “Standard of Electronic Information Products Environmental Friendly Use Period”...
Regulatory compliance information Indications for use The Coronis Fusion 6MP (MDCC-6230) display is a display device capable of displaying two 3MP display images electronically fused together on the one panel. It is intended to be used in displaying and viewing digital images for review and analysis by trained medical practitioners.
Page 46
8. Important information Emissions test Electromagnetic environment – Compliance Guidance RF emissions Group 1 The Coronis Fusion 6MP uses RF energy only for its internal CISPR 11 function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Page 47
8. Important information Immunity test Electromagnetic IEC 60601 Compliance level environment – Test levels guidance Voltage dips, short Mains power quality < 5% U (> 95% dip in < 5% U (> 95% dip in interruptions and voltage ) for 0.5 cycle ) for 0.5 cycle should by that of a typical variations on power...
Page 48
8. Important information Immunity test Electromagnetic IEC 60601 Compliance level environment – Test levels guidance than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with symbol: At 80 MHz and 800 MHz, the higher frequency range applies. These guidelines may not apply in all situations.
8. Important information These guidelines may not apply in all situations. Electromagnetic propagation is af- fected by absorption and reflection form structures, object and people. Explanation of symbols Symbols on the device On the device or power supply, you may find the following symbols (nonrestrictive list): Indicates compliance with the Directive 93/42/EEC as Class I device Indicates compliance with the Directive 93/42/EEC as Class II device Indicates compliance with Part 15 of the FCC rules (Class A or Class B)
Barco software products are the property of Barco. They are distributed under copyright by Barco N.V. or Barco, Inc., for use only under the specific terms of a software license agreement between Barco N.V. or 6. Values for xx and yy can be found in the technical specifications paragraph.
8. Important information Barco Inc. and the licensee. No other use, duplication, or disclosure of a Barco software product, in any form, is authorized. The specifications of Barco products are subject to change without notice. Trademarks All trademarks and registered trademarks are property of their respective owners.
Page 52
8. Important information Maximum luminance 720 cd/m² DICOM calibrated luminance 500 cd/m² (ULT off) Contrast ratio (typical) 1000:1 (panel typical) Response time (Tr + Tf) 18 ms Scanning frequency (H; V) 30-150 kHz; 15-80 Hz Housing color Black / Silver Video input signals DVI-D Dual Link (2x), DisplayPort (2x) Video inout terminals...
Open source license usage This product contains software components released under an Open Source license. A copy of the source code is available on request by contacting your Barco customer support representative. EACH SEPARATE OPEN SOURCE SOFTWARE COMPONENT AND ANY RELATED DOCUMENTA- TION ARE PROVIDED "AS IS"...
Need help?
Do you have a question about the MDCC-6230 and is the answer not in the manual?
Questions and answers