Page 1
NIV - Non-Invasive Ventilation Mask Ventilation Operating Instructions Addendum Evita 4 as of Software 4.n Evita 2 dura as of Software 4.n...
Page 2
N N N N O O O O T T T T I I I I C C C C E E E E P P P P r r r r o o o o p p p p r r r r i i i i e e e e t t t t a a a a r r r r y y y y I I I I n n n n f f f f o o o o r r r r m m m m a a a a t t t t i i i i o o o o n n n n This document contains information in which Dräger Medical, Inc.
Important Safety Information Operator's Responsibility for Patient Safety Important Safety Information Limitation of Liability Operator's Responsibility for Patient Safety Dräger Medical, Inc.'s liability, whether arising out of or related For correct and effective use of the product and in order to to manufacture and sale of the goods, their installation, avoid hazards it is mandatory to carefully read and to demonstration, sales representation, use, performance, or...
Important Safety Information Warranty Warranty Application of this warranty is subject to the following All Dräger products are guaranteed to be free of defects for a conditions: period of one year from date of delivery. The following are exceptions to this warranty: 1.
Important Safety Information Definitions Definitions WARNING ! A WARNING statement refers to conditions with a possibility of personal injury if disregarded. CAUTION ! A CAUTION statement designates the possibility of damage to equipment if disregarded. N N N N O O O O T T T T E E E E : : : : A NOTE provides additional information intended to avoid inconveniences during operation.
NIV adds the choice between mask ventilation and ventilation of intubated patients. The NIV option extends the range of applications for Evita 4 and Evita 2 dura ventilators as of software 4.n. WARNING ! The Evita 4 and Evita 2 dura intensive care ventilators must only be used under the supervision of qualified...
Installing the NIV-Option NIV - Non-invasive Ventilation Using NIV Installing the NIV-Option WARNING ! Installation of the NIV option should only be performed by factory trained and authorized service personnel follo- wing the respective installation documentation. Using NIV WARNING ! Deadspace will increase when using a mask.
The "Tube/Mask" menu will appear. Turn dial knob to select » Mask « screen key. Press dial knob to confirm. NIV will appear in the status display. The background color of the status page will switch from blue to green.
Using NIV Setting Ventilation Parameters Setting Ventilation Parameters Perform in the same fashion as in » Tube « application mode. For CPAP/PS, an additional screen dial » Tinsp « appears on screen. The ventilator limits the maximum duration of a pressure support breath to 4 seconds in »...
Using NIV Leak Compensation in »Mask« Application Mode The following alarm messages are not displayed in »Mask« application mode — P. supp. > 4 s — P. supp. > 1,5 s — P. supp. > Tinsp — Leakage WARNING ! Always check alarm limits when switching from »Mask«...
Message with medium priority Advisory Message with low priority In the table below, only messages specific to the NIV application mode are listed. This table is intended to help identifying possible causes of an alarm and to assist with prompt corrective action.
Ordering Information Leak Compensation in »Mask« Application Mode Ordering Information Name/Description Order No. Kit "Non-invasive Ventilation NIV" 84 14 474...
Page 16
These Instructions for Use apply only to Evita 4 or Evita 2 dura with Serial No.: If no Serial No. has been filled in by Dräger, these Instructions for Use are provided for general information only and are not intended for use with any specific machine or device.
Need help?
Do you have a question about the NIV and is the answer not in the manual?
Questions and answers