Siemens SITRANS F Series Operating Instructions Manual page 153

Coriolis flowmeters
Hide thumbs Also See for SITRANS F Series:
Table of Contents

Advertisement

Table B-13
Zero point adjustment
Modbus ad‐
Data type /
dress
Size [bytes]
2132
Unsigned /
2
2133
Float / 4
2135
Unsigned /
2
2136
Float / 4
2138
Float / 4
2140
Float / 4
2142
Float / 4
2144
Unsigned /
2
SITRANS FC310
Operating Instructions, 01/2019, A5E44036384-AB
Parameter
Description
Zero Point Adjust‐
Select zero-point adjust‐
ment
ment method. Automatic
zero point adjustment is rec‐
ommended.
● 0 = Auto
● 1 = Manual
Manual Zero Point
Enter agreed zero point off‐
Offset
set value for manual zero
point adjustment mode.
Zero Point Duration
Define duration of zero point
adjustment.
Standard Deviation
Standard deviation during
auto zero point adjustment
Standard Deviation
Set limit for zero point adjust‐
Limit
ment Standard Deviation val‐
ue. If the Standard Deviation
exceeds the Standard Devi‐
ation Limit, the auto zero
point adjustment is aborted.
Zero Point Offsett
Set limit for zero point offset.
Limit
If the zero point offset ex‐
ceeds the zero point offset
limit, the zero point offset
cannot be stored
Zero Point Offset Val‐
Default zero point offset
ue
based on factory calibration
of sensor.
A Zero point offset compen‐
sates for sensor variations
due to process conditions.
Zero Point Adjust
Shows the progress of the
Progress
currently running Zero Point
adjustment in percentage
Modbus communication
B.3 Modbus holding registers tables
Default value
Value range Access
[unit]
0
0 to 1
0 [kg/s]
0 to 1023
30 [s]
1 to 999
0 [kg/s]
-1023 to
+1023
Sensor size
0 to +1023
specific [kg/
s]
1)
Sensor size
-1023 to
specific [kg/
+1023
s]
1)
0 [kg/s]
-1023 to
+1023
0 [%]
0 to 100
level
Read /
Write
Read /
Write
Read /
Write
Read only
Read /
Write
Read /
Write
Read only
Read Only
153