Siemens SITRANS F Series Operating Instructions Manual page 149

Coriolis flowmeters
Hide thumbs Also See for SITRANS F Series:
Table of Contents

Advertisement

Modbus ad‐
Data type /
dress
Size [bytes]
2442
Float / 4
2444
Float / 4
Table B-8
Totalizer
Modbus ad‐
Data type /
dress
Size [bytes]
2609
unsigned / 2 Totalizer State
2610
float / 4
3018
Unsigned /
4
SITRANS FC310
Operating Instructions, 01/2019, A5E44036384-AB
Parameter
Description
Density Correction
Set density compensation
Factor
value (gain) in order to make
a density correction (scale
factor).
To increase the displayed
density value with +0.5 %,
set the density factor to
1.005.
The displayed density value
will now be 0.5 % higher
than before
Density Correction
Set density compensation
Offset
value (offset) in order to
make an offset on the meas‐
ured density.
To make the flowmeter show
+ 2 kg/m
ty offset to 2.000 kg/m
'Sensor' menu
Parameter
Description
Totalizer state
● 0 = paused
● 1= running
Totalizer Value
The totalized MASS value in
kg
Totalizer fixed
The totalized MASS value in
point part
kg. Most significant word of
the totalizer (MSW).
The format of the totalizer is
a TotalType.
The TotalType represent a
fixed point value in the MSW
32 bit and a fractional part in
the LSW 32 bit.
Example: 2.03 would be rep‐
resented as fixed point part =
2 and fractional part =
30000000
B.3 Modbus holding registers tables
Default value
[unit]
1
0 [kg/m
, change the densi‐
3
in the
3
Default val‐
ue [unit]
1
0 [kg]
0 [kg]
Modbus communication
Value range Access
level
-1.999 to
Read /
+1.999
Write
]
-1 400 to +1
Read /
3
400
Write
Value range
Access
level
0 to 1
Read only
Min. -1.70E+38
Read Only
Max. 1.70E+38
Min
Read Only
-2147483648
Max
2247483647
149