SITRANS FC310
Operating Instructions, 01/2019, A5E44036384-AB
Parameter
Description
Mass flow Simulation
Set mass flow simulation val‐
Value
ue.
The mass flow will be set to
this value on all outputs,
if Simulation Mass Flow is
enabled
Density Simulation
Set density simulation value.
Value
The density will be set to this
value on all outputs, if Simu‐
lation Density is enabled
Media Temperature
Set media temperature sim‐
Simulation Value
ulation value. The media
temperature will be set to
this value on all outputs
if Simulation media Temper‐
ature is enabled
Frame Temperature
Set frame temperature simu‐
Simulation Value
lation value.
The frame temperature will
be set to this value on all out‐
puts if Simulation Frame
Temperature is enabled
Volume flow Simula‐
Set volume flow simulation
tion Value
value.
The volume flow will be set
to this value on all outputs,
if Simulation Volume Flow is
enabled
Enable Simulation
Activate simulation.
Select one of the following
values:
● Bit 0: Mass flow
● Bit 1: Density
● Bit 2: Volume flow
● Bit 3: Media temperature
● Bit 4: Frame temperature
Modbus communication
B.3 Modbus holding registers tables
Default value
Value range Access
[unit]
0 [kg/s]
-1023 to
+1023
1000 [kg/m
]
-20000 to
3
+20000
0 [°C]
-50 to +200
0 [°C]
-50 to +200
m
/s
-65 to +65
3
0
0 to 63
level
Read /
Write
Read /
Write
Read /
Write
Read /
Write
Read /
Write
Read /
Write
157