UNIVERSAL GAS DETECTOR
INSTALLATION

4.1.15 Remotely Mounted Sensor Calibration Fitting

When remote mounting the Meridian sensor head(s) over long distances, or at much lower or higher elevations
than the Meridian transmitter, the sensors head(s) can be setup for remote calibration. Remote calibration fittings
(077-1385 and 077-1386) screw directly into the Meridian sensor end cap. Refer to
Figure 4-10: Remotely
Mounted Sensor with Calibration
Fitting.
1. Remove the screw on each end-cap.
2. Screw in the fitting.
3. Run the appropriate calibration tubing from the transmitter to the sensor head fitting.
Figure 4-10: Remotely Mounted Sensor with Calibration Fitting