7. Maintenance

7.1. Calibrating the Device

WARNING: THE DEVICE SHIPS WITH FACTORY-CALIBRATED SENSORS. SPARE SENSORS
DO NOT NOT ARRIVE PRE-CALIBRATED. ALWAYS CALIBRATE SPARE SENSORS BEFORE
USING. FAILURE TO DO SO COULD RESULT IN INJURY OR DEATH.
WARNING: OPERATING A DEVICE THAT HAS EXCEEDED ITS CALIBRATION DATE CAN
CAUSE FALSE READINGS OF DETECTED GASES. READINGS OBTAINED WHILE DEVICE IS
OUT OF CALIBRATION ARE INVALID AND COULD LEAD TO INJURY OR DEATH.
WARNING: DURING CALIBRATION, THE DEVICE WILL NOT DETECT HAZARDOUS GASES.
ENSURE YOU NOTIFY AFFECTED PERSONNEL WORKING IN THE AREA THAT IT WILL BE OUT
OF SERVICE AND PROVIDE OTHER GAS DETECTION PROTECTION IN ITS PLACE. FAILURE TO
DO SO COULD RESULT IN INJURY OR DEATH.
WARNING: BEFORE YOU BEGIN, READ AND UNDERSTAND THE MSDS AND WARNING
LABELS FOR THE CALIBRATION GASES. FAILURE TO DO SO MAY RESULT IN SERIOUS INJURY
OR DEATH.
WARNING: CALIBRATION MODE INHIBITS LOCAL ALARMS IN THE DEVICE (THAT IS, A
TIME DELAY SET TO PREVENT UNWANTED ALARM TRIPS). VERIFY THE ENVIRONMENT IS
CLEAN AND FREE OF HAZARDOUS GASES AND TOXINS OR HAVE MONITORING
PERFORMED BY ANOTHER DEVICE PRIOR TO CALIBRATING. FAILURE TO DO SO COULD
LEAD TO INJURY OR DEATH (SEE
NOTE: The Company recommends that users consult 60079-29-2 for additional information on cal-
ibration and selection of appropriate calibration gas.
Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is
Section 5.1.3 Configuring the Setup

USE AND DISCLOSURE OF DATA

prohibited.
MERIDIAN
UNIVERSAL GAS DETECTOR
MAINTENANCE
Menu).
7-1