6. Functional checks

6.1 Intervals

Carry out a functional check on the SOMNOcomfort 2e and the SOMNOaqua at least every
6 months.
If you find a defect during the functional check, do not use the unit until the defect has
been remedied.

6.2 Function check of SOMNOcomfort 2e

1.
Assemble SOMNOcomfort 2e with the breathing hose, exhalation system, mask and
power supply cable ready for operation.
2.
Press the on/off switch to switch on the unit. If softstart is switched on, switch it off
with the softstart button. The fan conveys air through the breathing hose to the mask
and the current pressure in hPa is shown in the display.
3.
Close the opening of the mask, e.g. using your thumb or hand.
Compare the pressure shown on the display with the CPAP pressure prescribed. There
should only be a divergence of max. ±1 hPa.
4.
Press the softstart button to switch on softstart. The softstart display appears and the
display shows the softstart time.
The pressure increases slowly and reached the CPAP pressure after the time selected.
The CPAP pressure is then shown on the display.
5.
Switch off the unit using the on/off switch.
If the valves or functions are not fulfilled, please send in the unit for repair to the dealer or
the manufacturer.