gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
15.2. Force amplifier installation
Date:
Time:
Seq.
Description
No.
10.
Verify that the mains plug is disconnected.
11.
Remove protective packaging carefully.
Note the amplifier serial number indicated on
the name plate
12.
Screw the amplifier to the brackets in front of
the motor compartment with Arsalis logo on
the top. Screws and washers should be
already placed on the brackets.
13.
Measure
impedance
enclosure (power connector backshell) to the
treadmill PE.
Iss.: A
Rev.: 9
Date: 07/12/2020
Test Step
Required/No
minal (incl.
tolerance)
Torque =
6Nm
of
the
amplifier
R < 1ohm
Page : 79
Name:
Measurements/Test Results
Measured/
Remarks/Calculations
Calculated
S/N:
4x DIN6798A-M6, Fabory
51460060001
4x ISO7380-M6x10, Fabory
07151060010
R=
Pass/
No
pass