7.7.1

[1.1] Input of basic settings

Procedure
1. "1 Quick Start" > "01 Basic settings" main menu
2. Call the parameter display:
Press <Enter>.
3. Set the display language.
If the language of the Quick Start Manual was also set as operating language, this is set
at the factory; otherwise the operating language is English.
Use the "Language" parameter field to select your preferred display language.
4. Setting parameters for date and time:
Note
Effects of changes
Changing the time has a direct effect on the sequence of logbook entries and all other
time-based device services.
• Check the logbook entries following changes to this setting. If necessary, reset the
– Specify display format of the date: "Date format" parameter field.
– Enter the current date: "Date" parameter field. Authorize access with Standard PIN, if
– Enter the current time: "Time" parameter field.
– Enable/disable automatic switching to daylight saving time: "Daylight saving time"
– Enter the date of commissioning: "Commissioning date" parameter field.
Quick Start
Compact Operating Instructions, 05/2018, A5E31805656-06
parameters of the logbook filter function.
This parameter field can be edited without PIN input. The following settings are
possible: DD.MM.YYYY, MM/DD/YYYY or YYYY-MM-DD.
necessary.
parameter field.
Select one of the options:
Setting
Description
Off
Automatic switching is disabled.
Autom. EU
Automatic switching to CEST ("Central European Summer Time") on the last
Sunday in March. The CEST ends on the last Sunday in October.
Autom. US
Switching to U.S. DST ("Daylight Saving Time") on the second Sunday in
March. DST ends on the first Sunday in November.
Each change results in a logbook entry that has to be acknowledged or confirmed.
Commissioning
7.7 Commissioning the device
87