Siemens SIPROCESS GA700 Quick Start Manual page 126

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Technical specifications
11.4 Technical specifications of analyzer modules
Influence variables
Sample gas pressure
Sample gas flow
Accompanying gases
Supply voltage
Table 11- 19 OXYMAT 7: Climatic conditions
Climatic conditions
Storage and transport
Max. permissible ambient temperature
surrounding the basic device during
operation
Ambient humidity (relative humidity)
during storage, transport or operation
Restriction for installing an ULTRAMAT 7 analyzer module: 5 to 45 °C
1)
Avoid condensation on the installed components!
2)
126
Deviation of the span gas
Deviation at zero point
Deviation of the span gas
Deviation at zero point
Deviation of the span gas
Analyzer module 24 V DC
(-5%/+10%)
Basic device
85 V AC to 264 V AC / 47 Hz to
63 Hz
1)
≤ 0.5% of the current measuring span/10 K or
≤ 50 vpm O
/10 K;
2
whichever is greater
≤ 0.2% of the smallest measuring span/1% pressure
change or
≤ 50 vpm O
/1% pressure change;
2
whichever is greater
Deviation ≤ 0.2% of current measuring span/1% pressure
change or
50 vpm O
/1% pressure change;
2
whichever is greater
≤ 1% of smallest measuring span per 0.1 l/min. change in
flow or
50 vpm O
per 0.1 l/min. change in flow within the permissi-
2
ble flow range (0.3 to 1 l/min);
whichever is greater
≤ 1% of current measuring span per 0.1 l/min. change in
flow or
50 vpm O
per 0.1 l/min. change in flow within the permissi-
2
ble flow range (0.3 to 1 l/min.);
whichever is greater
Zero point deviation (cross-sensitivity) in accordance with
Table A.1 of EN 61207-3
< 0.1% of the current span
< 0.1% of the current span
-30 to +70 °C
0 to 50 °C
< 90%
2)
Compact Operating Instructions, 05/2018, A5E31805656-06
Quick Start

Advertisement

Table of Contents
loading

Table of Contents