Siemens SIPROCESS GA700 Quick Start Manual page 122

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Technical specifications
11.3 Technical specifications of the wall housing
Table 11- 9
Electrical inputs and outputs
Digital outputs (DO)
Digital inputs (DI)
Ethernet interface (bottom)
Service interface (inside at PU)
Table 11- 10 Wall housing: Option modules
Option modules
Option module 1.1
Option module 2.1
Option module 2.2
Table 11- 11 Wall housing: Climatic conditions
Climatic conditions
Permissible operating altitude
Permissible ambient temperature dur-
ing storage and transport
Ambient humidity (rel. humidity) during
storage, transport or operation
Permissible ambient temperature and
humidity
122
Wall housing: Electrical inputs and outputs
8, with changeover contacts, parameters can be freely
assigned, e.g. for measuring range identification;
max. load: 24 V AC/DC/1.7 A (total load for all 8 relay out-
puts in continuous operation max. 160 W), floating, non-
sparking
8, designed for 24 V, isolated, freely parameterizable, e.g.
for autoranging
Ethernet RJ 45, 100 MBit
Ethernet RJ 45, 100 MBit
12 digital outputs, with changeover contacts, max. load: 24
V AC/DC/1.7 A (total load for all 12 relay outputs in continu-
ous operation max. 244 W), floating, non-sparking
8 digital inputs, designed for 24 V, isolated, freely parame-
terizable
6 analog outputs, 0/4 to 20 mA, floating; maximum load
750 Ω
6 digital outputs, max. load: 24 V AC/DC/1.7 A (total load
for all 6 relay outputs in continuous operation max. 122 W),
floating, non-sparking
6 analog outputs 0/4 to 20 mA, isolated; max. load 750 Ω
4 analog inputs 0/4 to 20 mA, non-isolated, internal re-
sistance 100 Ω
4 digital inputs, designed for 24 V, isolated
3 000 m above sea level
-30 ... +65 °C
< 90%
Technical specifications of analyzer modules (Page 123)