Siemens SIPROCESS GA700 Quick Start Manual page 24

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Combination operation
4.3 Gas connections
CALOMAT 7
Table 4- 3
Operating mode
Pressure operation/Suction operation
OXYMAT 7 / ULTRAMAT 7 / CALOMAT 7
Table 4- 4
Operating
mode
Pressure
operation
Suction ope-
ration
24
Mounting position for the sample gas restrictor: CALOMAT 7
Mounting position for the sample gas restrictor
OXYMAT 7 /
ULTRAMAT 7
OXYMAT 7 must always
be purged first.
Install a sample gas
restrictor in front of the
sample gas inlet of the
OXYMAT 7.
ULTRAMAT 7 must
always be purged first.
Install a sample gas
restrictor in the sample
gas outlet of the
OXYMAT 7.
Replace a sample gas
restrictor in the sample
gas inlet of the
OXYMAT 7 with a clamp-
ing screw.
Clamping screw
Operation with one or two CALOMAT 7
Pre-assembled clamping screws. No retrofitting or disman-
tling.
Sample gas restrictor/clamping screw
ULTRAMAT 7 / CALOMAT 7
We recommend purging the
CALOMAT 7 as the first
module
Compact Operating Instructions, 05/2018, A5E31805656-06
OXYMAT 7 / CALOMAT 7
No restrictions
Quick Start