Arrangement of the gas connections in the wall-mounted device
①
②
③
④
1
2
3
4
P
Figure 6-5
Design of the gas connections
Gas connection
Sample gas
Reference gas
Quick Start
Compact Operating Instructions, 05/2018, A5E31805656-06
Purging gas inlet
Purging gas outlet
Slot of analyzer module 2
Slot of analyzer module 1: ULTRAMAT 7
Sample gas inlet
Sample gas outlet
Reference gas outlet
Reference gas inlet
Atmospheric pressure sensor
ULTRAMAT 7 gas connections in the wall-mounted device, bottom
Material
Stainless steel Material No. 1.4404
or Hastelloy Material No. 2.4819
Stainless steel Material No. 1.4404
Connecting
6.1 Gas connections
Nozzle external diameter
6 mm
6 mm
47