Siemens SIPROCESS GA700 Quick Start Manual page 120

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Technical specifications
11.2 Technical specifications of rack-mounted housing
Electrical characteristics
EMC interference
immunity (electro-
magnetic compati-
bility) during
operation:
Electrical safety
Power unit cable
Table 11- 3
Electrical inputs and outputs
Digital outputs (DO)
Digital inputs (DI)
Service interface (rear)
Service interface (front)
Table 11- 4
Option modules
Option module 1.1
Option module 2.1
Option module 2.2
120
with one and/or two
OXYMAT 7 /
ULTRAMAT 7 units
with one or two
CALOMAT 7 units
Rack-mounted enclosure: Electrical inputs and outputs
Rack-mounted enclosure: Option modules
in accordance with the standard requirements of
NAMUR NE21 (05/2006) and EN 61326-1 (2013)
in accordance with the standard requirements
NAMUR NE21 (05/2006) and EN 61000-6-
2:2005
in accordance with EN 61010-1, overvoltage category II
Cable diameter: max. diam. 10 mm
Core cross-section: max. 3 x 2.1 mm
8, with changeover contacts, parameters can be freely
assigned, e.g. for measuring range identification; max. load:
24 V AC/DC/1.7 A (total load for all 8 relay outputs in con-
tinuous operation max. 160 W), floating, non-sparking
8, designed for 24 V, isolated, freely parameterizable, e.g.
for autoranging
Ethernet RJ 45, 100 Mbps
Ethernet RJ 45, 100 Mbps
12 digital outputs, with changeover contacts, max. load: 24 V
AC/DC / 1.7 A
(total load for all 12 relay outputs in continuous operation
max. 244 W), floating, non-sparking
8 digital inputs, designed for 24 V, floating, freely configurable
6 analog outputs, 0/4 to 20 mA, floating; maximum load:
R
≤ 750 Ω
L
6 digital outputs, load rating: 24 V AC/DC/1.7 A (total load for
all 6 relay outputs in continuous operation max. 122 W), float-
ing, non-sparking
6 analog outputs, 0/4 to 20 mA, floating; maximum load:
R
≤ 750 Ω
L
4 analog inputs 0/4 to 20 mA, non-isolated, internal re-
sistance ≤ 100 Ω
4 digital inputs, designed for 24 V, floating
Compact Operating Instructions, 05/2018, A5E31805656-06