Technical Specifications; Determining The Technical Specifications; Technical Specifications Of Rack-Mounted Housing - Siemens SIPROCESS GA700 Quick Start Manual

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Technical specifications

11.1

Determining the technical specifications

The technical specifications are obtained based on the specifications of DIN EN 61207-1.
Unless specified otherwise, the data listed below relates to the following measurement
conditions.
Technical specifications
Ambient temperature
Ambient pressure
Sample gas flow
Sample gas humidity
Site of installation
Reference gas
Reference application
For devices with filled reference gas side
1)
The technical specifications for time and measuring response as well as for the influencing variables can sometimes
2)
differ significantly for other gas mixtures
11.2

Technical specifications of rack-mounted housing

Table 11- 1
General
Installation position
Weight
Degree of protection
Table 11- 2
Electrical characteristics
Auxiliary power supply
Power consumption
Quick Start
Compact Operating Instructions, 05/2018, A5E31805656-06
OXYMAT 7
Free of vibration, shock and
impact
Nitrogen (N
)
2
-
Rack-mounted enclosure: General technical specifications
Rack-mounted enclosure: Electrical characteristics
ULTRAMAT 7
25 °C
Atmospheric (approx. 1 000 hPa)
0.6 Nl/min; Nl = Standard liter
Dew point < -40°C
Vibration- and
impact-free
Standard: Nitrogen (N
Option: freely selectable
-
Horizontal
Approx. 8.5 kg
IP20 according to EN 60529
100 to 240 V AC (nominal range of use 85 to 264 V)
50 to 60 Hz (nominal range of use 47 to 63 Hz)
Max. 280 VA
11
CALOMAT 7
Vibration- and
impact-free
)
-
2
1)
Hydrogen in nitrogen
(H
in N
)
2
2
2)
119

Advertisement

Table of Contents
loading

Table of Contents