Chapter 2 Performance Verification Test - Rigol DS1000Z-E Series Performance Verification Manual

Digital oscilloscope
Hide thumbs Also See for DS1000Z-E Series:
Table of Contents

Advertisement

Chapter 2 Performance Verification Test

Chapter 2 Performance Verification Test
This chapter introduces the performance verification test methods and procedures of DS1000Z-E
series digital oscilloscope by taking DS1202Z-E as an example. Fluke 9500B is used in this manual for
the tests. You can also use other devices that fulfill the "Specification" in Table 2-1.
Table 2-1 Test Devices Required
Device
Oscilloscope
Calibrator
50 Ω Impedance
Adapter
Note:
1.
Make sure that the oscilloscope passes the self-test and self-calibration is performed before
executing the performance verification tests.
2.
Make sure that the oscilloscope has been warmed up for at least 30 minutes before executing any
of the following tests.
3.
Please reset the instrument to the factory setting before or after executing any of the following
tests.
DS1000Z-E Performance Verification Guide
Specification
DC output voltage range:
1 MΩ: 1 mV to 200 V
50 Ω: 1 mV to 5 V
Fast edge signal rise time: ≤ 150 ps
BNC (M)-BNC (F) cable
Recommended
Model
Fluke 9500B
--
RIGOL
2-1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds1202z-eQ640392

Table of Contents