Rigol DS1000Z-E Series Performance Verification Manual page 26

Digital oscilloscope
Hide thumbs Also See for DS1000Z-E Series:
Table of Contents

Advertisement

RIGOL
Bandwidth Limit Test
Vertical
Channel
Scale
100 mV/div
CH1
500 mV/div
100 mV/div
CH2
500 mV/div
[1]
Note
: Amplitude Loss (dB) = 20 × lg (Vrms2/Vrms1).
Time Base Accuracy Test
Channel
Test Result ΔT
CH1
[1]
Note
: Calculation Result = Test Result ΔT/1 ms.
[2]
Note
: For the number of years that the instrument has been used, please calculate according to the date in the
verification certificate provided when the instrument leaves factory.
Zero Point Offset Test
Fast Edge Signal
Channel
Amplitude
1.2 Vpp
CH1
3 Vpp
1.2 Vpp
CH2
3 Vpp
2
Test Result
Vrms1
Vrms2
Calculation
[1]
Result
±(25 ppm + 5 ppm/year × number of years
Vertical Scale
200 mV/div
500 mV/div
200 mV/div
500 mV/div
Amplitude
Loss
[1]
-3 dB to 3 dB
Limit
that the instrument has been used
Test Result
≤ 2.5 ns
DS1000Z-E Performance Verification Guide
Appendix Test Record Form
Limit
Pass/Fail
Pass/Fail
[2]
)
Limit
Pass/Fail