200 mV/div, 500 mV/div, 1 V/div, 2 V/div, 5 V/div and 10 V/div respectively.
2)
Adjust the output voltage of Fluke 9500B to 3 × the current vertical scale and -3 × the
current vertical scale respectively.
3)
Repeat steps 3 to 7 and record the test results.
4)
Calculate the relative error of each vertical scale: |(Vavg1 - Vavg2) - (Vout1 - Vout2)|/
Full Scale × 100%.
10. Turn off CH1. Test the relative error of each scale of CH2 using the method above and record the
test results.

Test Record Form

Vertical
Channel
Scale
1 mV/div
2 mV/div
5 mV/div
10 mV/div
20 mV/div
50 mV/div
CH1
100 mV/div
200 mV/div
500 mV/div
1 V/div
2 V/div
5 V/div
10 V/div
1 mV/div
2 mV/div
5 mV/div
10 mV/div
20 mV/div
50 mV/div
CH2
100 mV/div
200 mV/div
500 mV/div
1 V/div
2 V/div
5 V/div
10 V/div
[1]
Note
: The calculation formula is |(Vavg1 - Vavg2) - (Vout1 - Vout2)|/Full Scale × 100%; wherein, Vout1 and Vout2
×
are 3
the current vertical scale and -3
2-4
Test Result
Vavg1
Vavg2
×
the current vertical scale respectively.
Chapter 2 Performance Verification Test
Calculation Result
[1]
DS1000Z-E Performance Verification Guide
Limit
Pass/Fail
≤ 4%
≤ 3%
≤ 4%
≤ 3%