8.
Rotate HORIZONTAL
Press MENU  Vrms at the left side of the screen of the oscilloscope to turn on the effective
9.
value measurement function. Read and record Vrms2.
10. Calculate the amplitude loss: Amplitude Loss A1 (dB) = 20 × lg (Vrms2/Vrms1) and
compare the result with the specification. At this point, the amplitude loss should be within the
specification range.
11. Keep the other settings of the oscilloscope in Step 3 unchanged and set the vertical scale to 500
mV/div.
12. Output a Sine with 1 MHz frequency and 3 Vpp amplitude via Fluke 9500B.
13. Repeat Step 5.
14. Output a Sine with 20 MHz frequency and 3 Vpp amplitude via Fluke 9500B.
15. Repeat Step 8-10.
16. Turn off CH1. Test CH2 using the method above.

Test Record Form

Vertical
Channel
Scale
100 mV/div
CH1
500 mV/div
100 mV/div
CH2
500 mV/div
[1]
Note
: Amplitude Loss (dB) = 20 × lg (Vrms2/Vrms1).
2-8
SCALE to set the horizontal time base to 50 ns/div.
Test Result
Vrms1
Vrms2
Chapter 2 Performance Verification Test
Amplitude
Limit
Loss
[1]
-3 dB to 3 dB
DS1000Z-E Performance Verification Guide
Pass/Fail