Appendix Test Record Form - Rigol DS1000Z-E Series Performance Verification Manual

Digital oscilloscope
Hide thumbs Also See for DS1000Z-E Series:
Table of Contents

Advertisement

Appendix Test Record Form

Appendix Test Record Form
Model:
Impedance Test
Channel
Vertical Scale
100 mV/div
CH1
500 mV/div
100 mV/div
CH2
500 mV/div
DC Gain Accuracy Test
Vertical
Channel
Scale
1 mV/div
2 mV/div
5 mV/div
10 mV/div
20 mV/div
50 mV/div
CH1
100 mV/div
200 mV/div
500 mV/div
1 V/div
2 V/div
5 V/div
10 V/div
1 mV/div
2 mV/div
5 mV/div
10 mV/div
20 mV/div
50 mV/div
CH2
100 mV/div
200 mV/div
500 mV/div
1 V/div
2 V/div
5 V/div
10 V/div
[1]
Note
: The calculation formula is |(Vavg1 - Vavg2) - (Vout1 - Vout2)|/Full Scale × 100%; wherein, Vout1 and Vout2
are 3 × the current vertical scale and -3 × the current vertical scale respectively.
Bandwidth Test
Vertical
Channel
Scale
100 mV/div
CH1
500 mV/div
100 mV/div
CH2
500 mV/div
[1]
Note
: Amplitude Loss (dB) = 20 × lg (Vrms2/Vrms1).
DS1000Z-E Performance Verification Guide
RIGOL DS1000Z-E Series Digital Oscilloscope
Performance Verification Test Record Form
Tested by:
Test Result
Test Result
Vavg1
Vavg2
Test Result
Vrms1
Vrms2
Test Date:
Limit
0.99 MΩ to 1.01 MΩ
Limit
Calculation
[1]
Result
≤ 4%
≤ 3%
≤ 4%
≤ 3%
Amplitude
Limit
[1]
Loss
-3 dB to 3 dB
RIGOL
Pass/Fail
Pass/Fail
Pass/Fail
1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds1202z-eQ640392