9.
Calculate the amplitude loss: Amplitude Loss (dB) = 20 × lg (Vrms2/Vrms1).
10. Keep the other settings of the oscilloscope in Step 3 unchanged and set the vertical scale to 500
mV/div.
11. Output a Sine with 1 MHz frequency and 3 Vpp amplitude via Fluke 9500B.
12. Repeat Step 5.
13. Output a Sine with 200 MHz frequency and 3 Vpp amplitude via Fluke 9500B.
14. Repeat Step 7-9.
15. Turn off CH1. Test CH2 using the method above and record the test results.

Test Record Form

Vertical
Channel
Scale
100 mV/div
CH1
500 mV/div
100 mV/div
CH2
500 mV/div
[1]
Note
: Amplitude Loss (dB) = 20 × lg (Vrms2/Vrms1).
2-6
Test Result
Vrms1
Vrms2
Chapter 2 Performance Verification Test
Limit
Amplitude
Loss
[1]
-3 dB to 3 dB
DS1000Z-E Performance Verification Guide
Pass/Fail