8. Ordering information and technical data
82
8.2.2 Development unit
Dimensions
Weight
Capacity
Agitation
Development
Number of ports
Temperature control
range
time
error
Development chamber
material
volume
Temperature sensor
Level sensor
Gel holder
8.2.3 Common data
Safety regulations
EMC
Safety regulation
Safety glass
Operation environment
room temp.
humidity
Electricity requirements
mains voltage
mains frequency
power
Power disturbance
Mains failure
300 x 300 x 138 mm (W x L x H)
4.8 kg
1 or 2 gels
Gels are rotated in solutions during development
Solutions are automatically pumped into and out of the
development chamber
9 ports are available for solutions to enter and exit the
chamber through; port 0 i reserved for waste.
Up to 50°C; the chamber only heats.
_ <4 min. to heat solutions from 20 to 45°C.
±4°C from programmed temperature once the programmed
temperature is reached.
Stainless steel
Approximately 70 ml of solution will be pumped into the
chamber.
Stainless steel.
Enclosed in glass.
Stainless steel.
This product meets the requirements of the EMC Directive
89/336/EEC through the harmonized standards EN 50082-
2 (emission) end EN 50082-1 (immunity).
Note: The declaration of conformity is valid for the
instruments when it is
• used in laboratory locations
• used in same state as it was delivered from
GE Healthcare Bio-Sciences AB except for alterations
described in the user manual.
• used as "stand alone" unit or connected to other CE
labelled GE Healthcare Bio-Sciences AB instruments or
other products as recommend.
This product meets the requirements of the low Voltage
Directive (LVD) 73/23/EEC through the harmonized
standard EN 61010-1.
Class 1 apparatus
4-40°C
Max. 95%.
100/120 V AC (120 model)
220-230/240 V AC (220 V model)
50-60 Hz
Maximum;
30 VA Separation and control unit
200 VA Development unit
330 VA together
The system is protected against mains disturbance and
static discharges.
For port failures lasting less than 5-10 seconds, running
methods will automatically continue when power i returned.
For power failures lasting more than 5-10 seconds, running
methods are set to pause, an alarm sounds, and a message
appears on the display informing you about the power
failure.