8.2 Technical data

8.2.1 Separation and control unit
A list of the technical data for PhastSystem instruments is given below.
Dimensions
Weight
Keyboard
Display
LEDs
Alarm
Capacity
Programs
Separation
Development
Programmable parameters
Separation
Development
Internal power supply
Circuitry protection
Voltage range
Error
Current range
Error
Power range
Error
Internal Vh integrator
range
Battery back-up memory
Separation compartment
Electrodes
number
material
width
spacing
Cooling/heating
Temperature
Sample application
Safety precautions
8. Ordering information and technical data
460 x 300 x 138 mm (W x L x H)
6.2 kg
31 tactile keys
40 digit alphanumeric liquid crystal display.
4 LEDs for status information
An audible alarm sounds at the end of
separation methods.
1 or 2 gels
9 methods available for programming;
each method contains 9 steps.
9 methods available for programming;
each method contains 9 steps.
Volt, current, power, temperature, duration
in volthours, sample application, and an extra
alarm.
Inlet port, outlet port, duration in minutes,
temperature, temperature compensation curve,
and an alarm.
Short circuit protection
10-2000 VDC
_ < 3% of actual value (-5 V for 10-2000 VDC)
0.1-50.0 mA
_ < 2% of actual value (-0.2 mA for 0.1-5.0 mA)
0.1-7.0 W
_ < 6% of actual value (-0.3 W for 0.1-1.0 W)
Integrates volts with time.
1-9999 volthours/step
Lithium battery, shelf life: 10 years
1 cathode and 2 anodes/gel
Platinized titanium
42 mm
37 mm; inner electrodes (IEF)
43 mm; outer electrodes
(native and SDS-PAGE)
Electronically cooled/heated by Peltier element
0-70°C
Automatic with sample applicators.
If the separation compartment lid is open
during a run, an alarm will sound and the
internal power supply is switched off.