Xylem MJK Oxix Installation And User Manual page 103

Dissolved oxygen transmitter
Table of Contents

Advertisement

Dissolved Oxygen Transmitter
Zero Point Calibration
Zero point Calibration
Oxygen
Set actual measurement
Back
Select
A zero point check or calibration is conducted as follows:
• The sensor is set for a zero point solution. The current measurement is
shown in line 2 (0.10mg/l in this example).
• Set the desired measuring value on the first line using the up and down
arrow keys.
• Save the current measurement setup by pressing OK.
The value is saved in the converter and will correct the measurement value of
the sensor.
See section Guide to Oxygen-free Solution for Checking Oxix D.O. Sen-
sor Zero Point.
GB Oxix Manual 1209