Dissolved oxygen transmitter
Table of Contents

Advertisement

Dissolved Oxygen Transmitter
Appendix A. Pop-up and Error Messages
Current overrange
Factory reset occurred
High alarm
Low alarm
mA connection
Calibration
Eeprom write error
Measuring start failed
Measuring timeout
Missing adjustment
Not in solid mode!
GB Oxix Manual 1209
Alarms
Alarm - The mA connection has exceeded
the 4 - 20 mA range. For example 3,5 or
20,5 mA.
Alarm - A factory reset has occured. Some
settings have changed. Check all used set-
tings.
Alarm - The high set point has been de-
tected.
Alarm - The low set point has been de-
tected.
Alarm - The mA connection is not connect-
ed. If the current-out is used, deactivate mA
Current out to stop this alarm.
Errors
Error - 1. Perform calibration (see page 21)
and/or check the units and ranges (see page
52).
Error - Eeprom write error. If error keeps
popping up, contact MJK.
Error - Start failed. Try again.
Error - Timeout. Try again.
Error - The measured value has not been
entered.
Error - Calibration can only be performed
when in solid mode. Change mode and try
again.
85
FW: 841013-009, 846001-002, 841513-009