Xylem MJK Oxix Installation And User Manual page 16

Dissolved oxygen transmitter
Table of Contents

Advertisement

Dissolved Oxygen Transmitter
Oxix
Sensor
®
Dissolved oxygen saturated
Measurement principle
Sensor check
Pressure
Materials
Cleaning system
Cable
Cable length
Response time
Flow velocity
Power supply
Output
Accuracy
Resolution
Temperature range
Temperature compensation
Enclosure rating
Dimensions
Mounting
Weight
Approvals
GB Oxix Manual 1209
0 - 30.0 mg/l or ppm (parts per million)
0 - 120 %
Optical, near infra-red fluorescence (l
Automatic self-diagnostics
6 bar (100 psi)
Boby and Head: PU, PVC, silicone, epoxy and stainless steel 316 SS
Optical lens: Sapphire
O-ring: Viton
Cable: PUR
Built-in air or water jet
Water supply: 35-50 psi (2,3 - 3,4 Bar)
Air supply: 40-60 psi (2,7 - 4 Bar)
4 x 0,326 mm
, shielded, outer diameter Ø 5.0 mm (22 AWG)
2
10 m (33 ft.), can be extended up to 600 mm
t
less than 1 second, t
90
95
No flow required
12 V DC (10 -16 V DC), approx. 10 mA
RS 485, 9600 baud, 2-wire
Better than 1% of actual reading (0,02 mg/l)
Better than 0,02 mg/l
0 - 60 ºC
0 - 50 ºC
IP 68 to IEC 529 (10 m), NEMA 6
48.5 x 130 mm (diameter x length)
1,5 inch NPT thread
0,25 inch NPT thread for jet spray cleaning tube
0,77 kg (1.7 lbs)
EN 61000-6-4:2001, EN 61000-6-2:2001
16
= 475 nm, I
in
out
less than 60 second
FW: 841013-009, 846001-002, 841513-009
= 600 nm)