Xylem MJK Oxix Installation And User Manual page 91

Dissolved oxygen transmitter
Table of Contents

Advertisement

Dissolved Oxygen Transmitter
B. Save Log Data
1. Select "LogData" in the menu bar and click "Save Log to File".
2. Assign a file name, find the destination you want to save the file to, and
click "Save". Consequently the file is saved, and the contents is displayed.
C. Upgrade Display Firmware
Important: Always upgrade the display firmware before you upgrade
the converter software.
1. Select "Firmware" in the menu bar and click "Display".
2. Locate the "MJK Field FW" directory on the CD-ROM, select the 841xxx-
xxx.hex file (or similar 841*.hex) and click "Open".
3. Click OK when the "Display Firmware" dialogue box appears and prompts
for programming the display Flash memory.
The display unit firmware is then transferred, and the display unit is conse-
quently programmed. Expect this process to take a few minutes.
GB Oxix Manual 1209
91
FW: 841013-009, 846001-002, 841513-009