12.6

Calibration timer

Calibration timer
Configuration item
Calibration interval
12.7

Display

General information
Configuration item
Language
Automatic logoff time
Display type
Temperature Unit
Sign. main value
Selection/setting op-
tion
0 to 9999 days
Selection/setting op-
tion
German
English
French
Spanish
0 to 15 min.
Normal
Large display
Bargraph
°C
°F
No signal
Temperature input
Compensation temp.
Uncomp. conductivity
Compens. Conductivity
Explanation
Time from one calibration to the next. The time at which a cali-
bration is due is indicated by the calibration alarm on device ver-
sions with a display.
In addition, the calibration alarm is displayed on the JUMO
AQUIS touch S/P.
Explanation
Selection of the operating language for the JUMO digiLineCR
The automatic logoff time can be set here. This time counts
down as soon as the user has logged in to the device. Once this
time elapses, the logged-in user is logged off automatically. If
the automatic logoff time is set to the value 0 s, then automatic
logoff is inactive. The user then remains logged in for the entire
length of the session.
 chapter 7.2.1 "Log-on/Log-out", page 43
In the normal and large display, 2 values are shown on the dis-
play in the measuring mode (main and secondary value).
In the bargraph display, the main value is shown as a numerical
value at the center in the measuring mode and visualized un-
derneath as a bargraph display. In contrast to the normal and
large display, the secondary value is not shown in this case. The
value range of the main value for the bargraph can be set (see
next table).
The compensated conductivity is the main value and the tem-
perature is the secondary value in the default setting. However,
this setting can also be changed to meet your requirements (see
further below in this table).
Temperature unit setting for the device
Signal source for the main value display
In the measuring mode, the main value is shown on the display
as the center value (the largest).
The appearance of the display can be set in the display type
(higher up in this table).
12 Configuration