11 Data overview
Data point
TAG number
Description
Sensor origin
11.3

Process values

Data point
Uncomp. actual value
Compensated actual value
Compensation temperature
interface
Conductivity invalid
Temperature
Temperature alarm signal
56
Explanation
Measuring point information
The "TAG number" identifies the measuring point with a unique ID assigned
by the user using the JUMO DSM software. With the aid of the "TAG num-
ber", the JUMO digiLine electronics can be assigned to a designated digital
sensor input of a JUMO AQUIS touch S/P. When "TAG check" is activated in
the JUMO digiLine master, matching of the sensor's "TAG number" and the
digital sensor input is checked by the JUMO AQUIS touch S/P. If there is no
match, the JUMO digiLine electronics are not linked to the master. The "TAG
number" of the JUMO digiLine electronics can be edited only with the JUMO
DSM software.
Text field for a description of the measuring point. The description can be ed-
ited only with the JUMO DSM software.
The "Sensor origin" is assigned by the JUMO digiLine master to which the
JUMO digiLine electronics were last connected and provides information
about the device on which the JUMO digiLine electronics were last connect-
ed.
Explanation
Electrolytic conductivity in the unit set in the configuration data and without
consideration of the effect of temperature
Measured value of the electrolytic conductivity in the unit set in the configu-
ration data and with correction for the effect of temperature
Temperature value received via the interface
This value is used for temperature compensation when the setting "Via inter-
face" was selected in the parameter "Compensation signal" in the configura-
tion.
JUMO digiLine Mode: The compensation temperature is transmitted from
the JUMO digiLine master.
Modbus Mode: The user must ensure that the "Compensation temperature
interface" is updated cyclically by the Modbus master (JUMO mTRON T).
The Modbus functionality is explained in detail in the separate Modbus de-
scription.
In the event of a fault during the conductivity measurement (e. g. as the result
of a measuring range violation or compensation error), this alarm is triggered
and generates a corresponding signal on the JUMO AQUIS touch S/P. It can
also be queried by a Modbus master (JUMO mTRON T).
only for conductivity sensors with an integrated temperature probe:
current value measured by the integrated temperature probe
In the event of a fault during the temperature measurement (e. g. as the re-
sult of a measuring range violation or compensation error), this alarm is trig-
gered and generates a corresponding signal on the JUMO AQUIS touch S/
P. It can also be queried by a Modbus master (JUMO mTRON T).