14

Technical data

14.1

RS485 interface for JUMO digiLine or Modbus RTU

Protocol
Device address
c
Data formats
Baud rates
adjustable minimum response time
a
The JUMO digiLine protocol assigns the interface parameters automatically during startup (Plug & Play).
b
The Modbus RTU protocol is used to operate the JUMO digiLine electronics on a JUMO mTRON T CPU. For op-
eration on a JUMO mTRON T, the interface parameters must be set prior to initial commissioning with the JUMO
DSM software.
c
Specification in useful bit - stop bit - parity format.
14.2

Analog inputs (sensor connection side)

Input for temperaure sensor
Measuring range
Pt100
Pt1000
Connection types
Measuring accuracy
Ambient temperature influence
Period
a
MR: measuring range span
Input for CR conductivity sensor
Units
a
Display ranges
Temperature compensation
a
JUMO digiLine
or Modbus RTU
1 to 247
8 - 1 - no parity
8 - 2 - no parity
8 - 1 - odd parity
8 - 1 - even parity
9600 baud
19200 baud
38400 baud
0 to 500 ms
-50 to +250 °C
-50 to +250 °C
2-wire/3-wire
a
±0.25 % of MR
0.1 % / K
500 ms
µS/cm
mS/cm
kΩ × cm
MΩ × cm
0.000 to 9.999
00.00 to 99.99
000.0 to 999.9
0000 to 9999
b
TC linear
for -50 to +100 °C and over 0.045 µS/cm
b
TC curve
for -20 to +150 °C and over 0.055 µS/cm
c
TDS
for -50 to +250 °C
Natural water DIN EN 27888 for 0 to 36 °C
Natural water with expanded range for 0 to 100 °C
ASTM D1125-95 neutral impurities for
0 to 100 °C and over 0.045 µS/cm
ASTM D1125-95 alkali and acid impurities for
0 to 100 °C and over 0.055 µS/cm
14 Technical data
b
67