9 Calibration
number of calibration logbook entries that can be saved on the PC. Canceled or failed calibrations (cal-
ibrations outside the admissible limits) are not saved in the logbook. Manual changes of calibration val-
ues are also documented. The following data are retained in the logbook:
Date and time
Calibration values determined or entered
Calibration mode (true calibration/manual entry of calibration values)
Calibration assessment (assessment of the calibration values determined during the true calibration)
Sensor replacement counter reading (to assign the calibration logbook entries to the individual sen-
sors in the sensor replacement history of the JUMO digiLine electronics)
The calibration logbook can be viewed on the JUMO AQUIS touch S/P and on the PC using the JUMO
DSM software.
9.7

Evaluation criteria for calibration

Calibration
Invalid
value [unit]
Relative cell
...
constant [%]
Calibration
Invalid
value [unit]
Temperature
...
coefficient
[%/K]
52
Evaluation criteria for calibration of relative cell constant
Warning
<
50
...
Evaluation criteria for calibration of temperature coefficient
Warning
OK
<
75 to 125
OK
<
0 to 8
Warning
<
...
≤ 150 <
Warning
<
Invalid
...
Invalid
...