Digital Interface - JUMO digiLine CR Operating Manual

Intelligent electronics with rs485 interface for connecting conductive conductivity sensors to jumo digiline systems
Table of Contents

Advertisement

12 Configuration
Configuration item
CIP temperature
SIP temperature
CIP cycle duration
SIP cycle duration
Maximum number of
CIP cycles
Maximum number of
SIP cycles
a
The counters for CIP and SIP cycles are automatically incremented by the JUMO digiLine electronics each time a
CIP or SIP process is recognized on the basis of the configured CIP/SIP temperatures and duration of the CIP/SIP
cycle.
12.5

Digital interface

NOTE!
The settings for the digital interface are obtained automatically in the JUMO digiLine mode and may then
no longer be changed.
Ensure that interface settings are not changed inadvertently when using the JUMO DSM software.
For operation on the JUMO mTRON T, the settings must be made in advance with the JUMO DSM soft-
ware.
Configuration item
Baud rate
Data format
Floating Point
Hardware address
Device address
Min. response time
62
Selection/setting op-
tion
-20 to+150 °C
0 to 9999 s
0 to 999
Selection/setting op-
tion
9600
19200
38400
8 - 1 - no parity
8 - 1 - odd parity
8 - 1 - even parity
8 - 2 - no parity
Standard
IEEE754_LITTLE
IEEE754_BIG
-
1 to 247
0 to 500 ms
Explanation
Temperature thresholds for identifying CIP/SIP cycles
If the CIP/SIP cycle takes place above one of these values for
the set duration of a CIP/SIP cycle, the values identify a suc-
cessfully completed CIP/SIP cycle and the CIP or SIP counter
is incremented. The respective counter is reset only after the
value has dropped below the CIP/SIP temperature.
Duration of a CIP/SIP cycle
Specifies the number of CIP/SIP cycles at which the CIP/SIP
alarm is triggered on the master
Explanation
Transmission speed (symbol rate) of the RS485 interface
Transmission format of the RS485 interface
Selectable transmission format for float values (floating-point
numbers)
non-configurable, permanently assigned address of the digiLine
electronics for unique identification (required for plug & play)
Bus user identification for the digiLine electronics
Minimum time from receipt of a query to sending of the re-
sponse
This parameter is used to adjust the response speed of the dig-
iLine electronics to slower bus users.
a

Advertisement

Table of Contents
loading

This manual is also suitable for: