Operation; General Information - JUMO digiLine CR Operating Manual

Intelligent electronics with rs485 interface for connecting conductive conductivity sensors to jumo digiline systems
Table of Contents

Advertisement

7

Operation

7.1

General Information

Operating the JUMO digiLineCR
The JUMO digiLineCRis available in device versions with or without a local membrane keyboard and dis-
play (see "JUMO digiLine device frontsCR", page 13).
The device version without local operation has a status LED on the device front that indicates the
operating status of the device:
Flashing green in one-second interval: Measuring mode
Flashing red in one-second interval: Error
Flashing red very quickly (5× per second): Serious error
Device versions with display and operating panel show their measured values and operating status
(e. g. error status) locally and allow local access to some settings, device information and calibration
functions of the device. There are 4 operating keys on the membrane keyboard:
Explanation
"OK"
Opens submenus and confirms entries
"Back"
Returns to the previous menu level; leaves Settings and discards entries and set-
ting changes
"Up"
Move cursor up in the current menu level, scroll up or adjust setting values higher
When the "Up" operating key is held down while adjusting numerical values, the
change in the value accelerates.
"Down"
Move cursor down in the current menu level, scroll down or adjust setting values
lower
When the "Down" operating key is held down while adjusting numerical values, the
change in the value accelerates.
Configuration and calibration via interfaces
All device versions can be connected to a PC via USB, and configured and calibrated using the JUMO
DSM software. Refer to the operating manual for the JUMO DSM software for more details.
The device versions with RS485 interface for JUMO digiLine can also be set and calibrated on the
JUMO AQUIS touch S/P in the corresponding menus for digital sensors. The JUMO mTRON T can also
access the data from the JUMO digiLine via ModbusCRModbus. The options for access to configure and
calibrate depend on the PLC implementation in the JUMO mTRON T (see interface description for the
Modbus on the JUMO digiLineCRand the documentation for the JUMO mTRON T).
Measuring mode
After the JUMO digiLine has been switched onCRthe JUMO logo appears briefly on the display in device
versions with a display. The device then switches to the measuring mode. The measured values for elec-
trolytic conductivity and temperature are displayed here. The "Up" and "Down" operating keys can be
used to scroll through 3 different display screens:
Main display screen with electrolytic conductivity and temperature
Detailed measurement display with compensated and uncompensated measured values
Bar graph display of the sensor stress level (for degree of current sensor stress,
seechapter 12.4 "Sensor monitoring", page 61)
7 Operation
Operating key
41

Advertisement

Table of Contents
loading

This manual is also suitable for:

202764202765

Table of Contents