Analog
The 0-10V range of the analog programming interface must be calibrated separately.
Follow the procedure exactly as for calibrating the 0-5V range, using the
Programming
corresponding 10V menu options. All the analog signals will be scaled by a factor of
Interface
2.
0-10V Range

Remote Interface Calibration Procedure

Calibration can also be done via front panel. See "Front Panel Calibration
Procedure" on page 183.
Calibration of voltage programming and readback are combined in a single
procedure.
Output
1. Set the load to open circuit. Attach a DVM across the output terminals.
Voltage
2. Minimum calibration level
3. Enter voltage data
4. Maximum calibration level
5. Enter voltage data
6. Power supply calculates and stores calibration constants.
Output
1. Set the power supply and load operate at full output. You must ensure the power
Current
2. Minimum calibration level
Release 3.0
Set the output voltage to 10% by sending the command:
CAL:OUTP:VOLT:LEV MIN
Enter the voltage read from the external DVM.
CAL:OUTP:VOLT:DATA <voltage>
Set the output voltage to 90% by sending the command:
CAL:OUTP:VOLT:LEV MAX
Enter the voltage read from the external DVM.
CAL:OUTP:VOLT:DATA <voltage>
supply is operating in current mode during current calibration. Place a shunt on
the load line so that you can measure the current. Attach a DVM across the shunt
Set the output current to 10% by sending the command:
CAL:OUTP:CURR:LEV MIN
Remote Interface Calibration Procedure
Calibration
187