Environmental Specifications (cont'd.)
Item
Atmospheric Pressure
Disposal

Safety

Item
Certification
Type of Protection
Against Electrical
Shock
Degree of Protection
Against Ingress of
Liquids
Handling of Disposable
Supplies and Other
Consumables
Patient Mode of
Operation
Patient Leakage
Current
Degree of Protection
Against Electrical
Shock
2046275-019H
Description
Storage 15% to 95% RH non-condensing
Operating 700 to 1060 hPa
Storage 500 to 1060 hPa
Batteries Disposing of battery by fire or burning will cause the battery to
explode. The battery is recyclable. Follow local environmental
guidelines concerning disposal and recycling. Batteries may be
returned to GE Healthcare for recycling.
Device Recyclable
Description
• UL 60601-1 classified
• UL classified for CAN/CSA C22.2 No. 601.1
• CB certified for IEC 60601-1
• Meets applicable AAMI EC-11 requirements
Class 1, internally powered
Ordinary
Use only parts and accessories manufactured or recommended
by GE Medical Systems Information Technologies. Follow
manufacturer's instructions for use for disposable/consumable
products. Follow local environmental guidelines concerning the
disposal of hazardous materials.
Continuous
<10 µA
Type BF defibrillation protection for the patient cable (acquisition
module)
MAC™ 3500
Technical Specifications
139