Maintenance
Operational Checks
1.
Power-up self-test passed?
2.
Simulated printed rhythm strip successful?
3.
Simulated printed ECG successful?
4.
Simulated ECG stored on media successful?
5.
Simulated ECG data transmitted successfully?
6.
Exercise device successfully communicates with device?
7.
Blood pressure monitor successfully communicates with device?
8.
Barcode reader successfully communicates with device?
9.
Trolley casters lock successfully?
Diagnostic Tests
1.
Battery Status Test meets Battery Current expectation?
2.
Display diagnostic test successful?
3.
Speaker test successful?
4.
Keyboard test successful?
5.
Writer diagnostic tests successful?
6.
Read/Write Floppy Diagnostic Test successful? (This test does not need to be
performed as the MAC3500 does not have a floppy drive.)
7.
External/Internal Modem Test successful?
8.
SD Card Test successful?
9.
Ethernet Module Test successful?
104