GE MAC 3500 Service Manual page 87

Resting ecg analysis system
Hide thumbs Also See for MAC 3500:
Table of Contents

Advertisement

3.
Remove the two fasteners holding the two ground wires on either side of the
display.
4.
Pivot up the display to access the cable connections on the main PCB.
NOTE:
5.
Disconnect the following cables from the main PCB board:
• Blue ribbon cable
• Display light cable
6.
Push back on each of the tabs holding the display assembly to the pivot, as
shown in the following figure, and gently pull the display assembly free.
7.
Reverse step
2046275-019H
If a KISS pump is installed, it must be removed to access the display panel
connectors from the main board. See
on page
89.
1
through step
MAC™ 3500
"Replacing the Optional KISS Pump"
6
to attach a new display assembly to the unit.
Maintenance
87