GE MAC 3500 Service Manual page 97

Resting ecg analysis system
Hide thumbs Also See for MAC 3500:
Table of Contents

Advertisement

This is the number that was used when the option codes for this system were
generated. The number entered here must match the serial number on the system
label.
• Printhead resistance
This number can be found on the printhead label.
• System language
Verify with the customer what language to select.
1.
From the Main Menu, select System Setup.
The System Setup menu opens.
2.
Press Shift + F2.
You are prompted for the service password.
3.
Type the service password and press Enter.
The Service-Only Setup window
opens.
4.
Enter the system's serial number.
5.
Verify that Update Primary Boot is set to No.
6.
Enter the Print head resistance.
7.
Select the appropriate language for the Keyboard.
8.
Select Return.
Restoring the System Setup
Use the following procedure to restore the system setup.
NOTE:
Before beginning this procedure, you must have the SD card on which you backed
up the system setup before removing the CPU board. See
Board" on page
1.
Power up the cart, if necessary.
2.
From the Main Menu, select System Setup.
3.
Enter the system password and press Enter.
4.
Select Restore Setup.
5.
Select From SD Card.
2046275-019H
91.
MAC™ 3500
Maintenance
"Replacing the CPU
97