Sony Tektronix WFM 1125 Service Manual page 95

Digital television waveform monitor option 0a / 0b / 0c / 0d
Hide thumbs Also See for Tektronix WFM 1125:
Table of Contents

Advertisement

VGA output
Finish
WFM 1125 Option 0A/0B/0C/0D Service Manual
Horizontal sweep:
Trigger mode:
. . . . . . . . . . . . .
Trigger coupling:
Trigger level:
. . . . . . . . . . . . .
Trigger slope:
. . . . . . . . . . . . .
Display mode:
. . . . . . . . . . . . .
Procedure.
1. Confirm the output voltage from MONITOR connector is 800 mV " 10%.
This test confirms that the VGA signal is output correctly.
The following equipment is required.
Equipment Required
1. Connect the VGA monitor to the VGA OUTPUT on the waveform monitor
rear panel using the VGA video cable.
2. Confirm the same view as waveform monitor is displayed on the monitor.
1. Press and hold the INTEN/MENU button and select FACTORY :
EXECUTE ! OK.
. . . . . . . . . .
200 ps/div
Auto
. . . . . . . . . .
AC
0 V
Positive
Color Grading or Infinite
VGA monitor
VGA video cable
Performance Verification
4-25