Resetting the waveform
Monitor
WFM 1125 Option 0A/0B/0C/0D Service Manual
Case 2: Option 0A
Type:
cd 1125 <Return>
load0A <Return>
Case 3: Option 0B
Type:
cd 750 <Return>
load0B <Return>
Following menu appears on the screen:
Select Download item:
0: ALL
1: Firmware
2: TTF Files
3: GF Files
4: XY files
5: Graticule Data
Q: quit
Item Number ?
Type:
0 <Return>
The download is completed in about 20 min for Option 0A and Option 0B, or
30 min for Option 0C and Option 0D. After the download has been finished,
the download program exit automatically.
Press and hold the INTEN/MENU button on the front panel to the Configure
menu. Turn the SELECT knob to select the FACTORY : EXECUTE and press
OK bezel button to reset factory settings.
NOTE. Be sure to reset the waveform monitor to factory settings after the
download process is completed. If you skip this procedure, the waveform monitor
may not perform correctly.