Sony Tektronix WFM 1125 Service Manual page 84

Digital television waveform monitor option 0a / 0b / 0c / 0d
Hide thumbs Also See for Tektronix WFM 1125:
Table of Contents

Advertisement

Performance Verification
4-14
Trigger level:
. . . . . . . . . . . . .
Procedure.
1. Align the 0 mV level of the signal with the center graticule on the oscillo-
scope.
2. Change the oscilloscope vertical offset to 700mV.
3. Confirm the signal level is 700 mV" 7 mV.
4. Change the oscilloscope waveform to CH1 off, waveform to CH2, and
vertical offset to –350 mV.
5. Align the –350 mV level of the signal with the center graticule on the
oscilloscope.
6. Change the oscilloscope vertical offset to 350 mV.
7. Confirm the signal level is 350 mV" 7 mV.
8. Press the TIMING SIGNALS button on the TSG1001 front panel to set the
TIMING SIGNALS to BOWTIE.
9. Change the oscilloscope settings to:
Waveform:
. . . . . . . . . . . . . . .
Vertical scale:
. . . . . . . . . . . . .
Vertical offset:
Horizontal scale:
Record length :
10. Press the MATH REF button on the oscilloscope front panel, and then press
the main and side menu buttons in the following order to create a Math
waveform:
Change Math waveform definition ³ Dual Wfm Math ³ Set 1st Source to
[Ch1] ³ Set operator to [–] ³ Set 2nd Source to [Ch2]
11. Change the oscilloscope settings to:
Waveform:
. . . . . . . . . . . . . . .
Zoom:
. . . . . . . . . . . . . . . . . . .
Vertical scale:
. . . . . . . . . . . . .
Horizontal scale:
–140 mV
CH1 and CH2
100 mV/div for CH1
100 mV/div for CH2
. . . . . . . . . . . .
0 V for CH1 and CH2
1 ms/div
. . . . . . . . . . .
. . . . . . . . . . . .
5000 points
CH1 and CH2 to Off
On
100 mV/div (5.0 X Vert) for Math1
5 ms/div (0.2X Horz)
. . . . . . . . . . .
WFM 1125 Option 0A/0B/0C/0D Service Manual