8-3. SelfCalibration Problems

Problem
1. The built-in calibration sensor
does not come out/stays out.
2. SelfCalibration cannot be
executed.
3. SelfCalibration failure
4. SelfCalibration is canceled while
in progress.
Error Code Table
If an error related to any of the following values occurs, the error code and error message are displayed
in the "Color" menu.
• Calibration target value
• Adjustable brightness settings
• Adjustable black level settings
Error Code
000020
000021
010141
****52
60
Chapter 8 Troubleshooting
Possible cause and remedy
• Turn off the main power, and then turn it on again a few minutes later.
• Check whether a color mode for executing SelfCalibration has been
set (see
"Mode Settings" (page
• Check whether the date and time are set correctly on the monitor (see
"Clock Adjustment" (page
• Check whether the execution schedule been set (see
(page
36)).
• Check whether the calibration targets are set correctly (see
Settings" (page
24)).
• Try calibrating the monitor by using ColorNavigator 7.
Refer to the error code table. If an error code that does not
appear in the error code table is displayed, contact your local
EIZO representative.
• If there is a change in the video signal from the external device
during SelfCalibration (the signal disappears, a signal is input during
a no-signal state, or other), SelfCalibration is canceled.
• Prevent any changes in the video signal while executing
SelfCalibration.
• If a scheduled SelfCalibration is canceled, it is executed again when
the monitor transitions to the power saving mode after an hour or
more or when the monitor is turned off using
also be executed regardless of the schedule (see
(page
38)).
The sensor failed to open.
Check whether there is any foreign object near the sensor.
Invalid target value has been set.
Check the target value.
The target black level is too low.
Up the target black level or set "Min".