7-4. Built-in calibration sensor and
SelfCalibration problems
Problem
1. The built-in calibration sensor does not
come out/stays out.
2. SelfCalibration cannot be executed.
3. SelfCalibration failure
Error Code Table
Error Code
0011
0013
0014
0030
0032
0033
0034
0036
0037
0038
0039
0040
0041
0050
0060
0061
• Calibration cannot be performed because the target brightness is too low. Increase the
target brightness, and then execute SelfCalibration again.
• There may be a problem with the target value setting. Recheck the target value of the
Color Gamut.
• It is possible that measurement could not be performed correctly.
- Ensure that the built-in calibration sensor is not exposed to strong light such as
direct sunlight.
- Try executing correlation using ColorNavigator 6.
• The target brightness may be too high. Lower the target brightness, and then execute
SelfCalibration again.
• Calibration may have failed. Ensure that the built-in calibration sensor is not exposed to
strong light such as direct sunlight, and then execute SelfCalibration again.
• The built-in calibration sensor may have failed. Contact your local EIZO representative.
• It is possible that measurement could not be performed correctly.
- Try executing correlation using ColorNavigator 6.
• Measurement failed. Execute SelfCalibration again.
• There may be a problem with the target value setting. Recheck the target value of the
Color Gamut.
• It is possible that measurement could not be performed correctly.
- Ensure that the built-in calibration sensor is not exposed to strong light such as direct
sunlight.
- Try executing correlation using ColorNavigator 6.
• It is possible that measurement could not be performed correctly.
- Try executing correlation using ColorNavigator 6.
• The built-in calibration sensor may have failed. Contact your local EIZO representative
• The built-in calibration sensor failed to open/close.
• Turn off the main power on the rear side of the monitor, wait for a few minutes before
turning the power back on, and then execute SelfCalibration again.
Possible cause and remedy
• Turn off the main power on the rear side of the monitor,
wait for a few minutes before turning the power back on,
and then execute SelfCalibration again.
• Check whether the date and time are set correctly
(see
"4-2. Setting Monitor Date and Time" (page
• Check whether the calibration schedule has been set
(see
"4-3. Setting Calibration Schedule" (page
• Check whether the calibration targets are set correctly
(see
"4-4. Setting the calibration targets" (page
• Set the SelfCalibration target using ColorNavigator 6.
• Refer to the error code table.
Possible cause and remedy
35)).
36)).
38)).
Chapter 7 Troubleshooting
49