Reflex Test Screen - Interacoustics AA222 Instructions For Use Manual

Impedance audiometer
Hide thumbs Also See for AA222:
Table of Contents

Advertisement

AA222 Instructions for Use – US

3.6.5 Reflex test screen

The upper bar of the soft keys indicates the function in auto mode, while the lower bar shows the function for
the soft keys in manual mode.
Function Key
Description
O
Activating manual mode in the reflex test allows single reflex measurements at
a time and optionally the pressure at which the reflex is measured can be
manually set (see T).
P
The number indicates the reflex activator intensity of the currently selected
reflex measurement (Q).
Q
The pressure slider gives an indication of which pressure reflex measurements
are set to be tested (in manual mode (O) only). The slider is moved by holding
the pressure button (see T) and turning the wheel.
R
The compliance meter gives an indication of the current non- compensated
compliance value and can be used to help set the pressure at peak pressure
or at an offset from peak pressure (in manual mode (O) only).
S
The currently selected reflex measurement is indicated by the prominent
rectangle around it. Within the reflex graph the numerical deflection value is
also displayed.
T
Pressing Pressure allows setting the pressure manually (see Q) (in manual
mode (O) only).
Press Exclude to exclude the highlighted test. Once excluded, press Include to
reinstate it as part of the measure.
U
Pressing the arrow up button moves the reflex selection to the previous reflex
row. Moving the selection sideways is done with the wheel (19).
Page 31