Indications for Use
!
Cautions (Continued)
Do not apply sensor over open wound, incision, or compromised skin. Inspect the sensor
site prior to applying the sensor.
Inspect the sensor application site in accordance with the sensor instructions for use to
ensure correct sensor alignment and skin integrity. Patient sensitivity to the sensor may vary
due to medical status or skin condition.
Do not autoclave, sterilize, immerse, or spray this device with liquid or use caustic or
abrasive cleaning agents. Do not use cleaning agents or cleaning products that contain
ammonium chloride.
Follow local, state and national governing ordinances and recycling instructions regarding
disposal or recycling of the device and device components, including batteries.
In compliance with the European Directive on Waste Electrical and Electronic Equipment
(WEEE) 2002/96/EC, do not dispose of this product as unsorted municipal waste. This
device contains WEEE materials; please contact your distributor regarding take-back or
recycling of the device. If you are unsure how to reach your distributor, please call Nonin for
your distributor's contact information.
A functional tester can only be used to assess the performance of the SpO
-only sensors.
2
If this device fails to respond as described, discontinue use until the situation is corrected by
qualified technical professionals.
For measurement, the device may not work when circulation is reduced. Warm or rub the
finger, or reposition the device.
In some circumstances, the device may interpret motion as good pulse quality during
measurement. Minimize patient motion as much as possible.
Do not simultaneously touch the accessible connector pins and the patient.
The device has been designed for use within the specified accuracy ranges. Use outside of
these ranges has not been tested and may result in improper oximeter performance.
All parts and accessories connected to the serial port of this device must be certified
according to at least IEC Standard EN 60950, IEC 62368-1, or UL 1950 for data processing
equipment.
Ensure time and date value is correct.
Implants may affect performance.
Refer to the applicable sensor Instructions for Use (IFU) for additional warnings and
cautions.