6-3 SETTING BCD OUTPUT

The number of output times, data logic, and EOC logic are set with the function selecting mode 'F-11.'
Since the hold input and output prohibit input are set to negative logic, no change is allowed.
'F-11' indication: (1) Output times – (2) Data logic – (3) EOC logic
(1) Setting the Number of Output Times
The first numeric value (1) indicated with the function selecting mode 'F-11' describes the following number
of output times. Select the desired item with arrow keys. See "4-1 OPERATION & FUNCTION."
0: 125 times/sec
1: 62.5 times/sec
2: 31.3 times/sec
3: 15.6 times/sec
(2) Setting Data Logic
The center value (2) indicated with the function selecting mode 'F-11' describes the following BCD output
logic. Select either of the two with the arrow keys.
In addition, when the signal is enabled and when the transistor is ON, it is considered as negative logic.
0: Negative logic
1: Positive logic
(3) Setting EOC Logic
The last numeric (3) indicated with the function selecting mode 'F-11' describes the following EOC logic.
Select either of the two with the arrow keys.
0: Negative logic
1: Positive logic

6-4 CONNECTING BCD OUTPUT

For connecting to external devices, separately prepare the following connector.
For connection, see "6-2 PIN NOS & SIGNAL NAMES" and the following figures.
(One example of the connection)
Connector in use: 57-30360 (DDK Ltd. equivalent)
Internal
Internal
 
 
 
 
 
Output Connection Diagra m
Output Connection Diagra m
Max 30V
Max 30V
External
External
← Max 20mA
← Max 20mA
COM
COM
12V
12V
Internal
Internal
1〜 5mA  →
1〜 5mA  →
Input Connection Diagra m
Input Connection Diagra m
31
Exte rnal
Exte rnal
Switch or
Switch or
Open Collector
Open Collector